cobas® 4000 Specifications for the cobas c 311 and cobas e 411 analyzers |
|
 |
| System |
 |
 |
cobas® 4000 analyzer series includes two stand-alone analyzers for serum work area
testing in small workload laboratories:
| • |
cobas c 311 is a fully automated, random access system for clinical chemistry and homogeneous immunoassay analysis. |
| • |
cobas e 411 is a fully automated, random access system for heterogeneous immunoassay analysis. It is available as both a disk system and a rack system. |
|
 |
| Test throughput |
 |
 |
Up to 300 tests/hr. photometric and ISE on cobas c 311
Up to 86 tests/hr. on cobas e 411 |
 |
| Type of analyzers |
 |
 |
cobas c 311: photometric measuring unit including ISE
cobas e 411: ECL technology measuring unit |
 |
| Number of channels |
 |
 |
Up to 63 channels (reagent slots) in total:
| • |
Up to 45 channels on cobas c 311 |
| • |
Up to 18 channels on cobas e 411 |
|
 |
| Available parameters |
 |
 |
196 parameters in total:
| • |
117 photometric tests, 3 ISE tests, 8 formulas, 3 serum indices on cobas c 311 |
| • |
65 heterogeneous tests with cobas e 411 |
|
 |
| Sample types |
 |
 |
Serum, Plasma, Urine, CSF, Whole Blood (HbA1c only) |
 |
| Sample container types |
 |
 |
| Primary tubes: |
5-10mL; 16x100, 16x75, 13x100, 13x75mm |
| Sample cup: |
2.5 mL |
| Micro cup: |
cobas c 311 only: 1.5mL |
| Cups on tube: |
Cup on top of a 16x75/100mm tube
cobas c 311 only: cup on top of one non-standard tube and
false buttom tube |
|
 |
| Result integrity |
 |
 |
| • |
Clot detection available on both analyzers |
| • |
Disposable tips on cobas e 411 |
| • |
3 test-specific serum indices on cobas c 311; automatic interference flagging of
results provided by cobas IT 3000 application |
|
 |
| Rerun/reflex function |
 |
 |
Automatic and manual rerun on cobas c 311, manual rerun only on cobas e 411
Automatic reflex is supported by both systems, reflex request to be provided by
cobas IT 3000 work area manager or LIS |
 |
| Regulatory requirements |
 |
 |
CE, UL, C-UL, CB-report and certificate |
 |
 |
 |
 |
cobas c 311 analyzer - Specifications |
|
 |
| System |
 |
 |
Fully automated, random access analyzer for Clinical Chemistry and Homogeneous Immunology
(HIA). |
 |
| Sample throughput |
 |
 |
Up to 300 samples/hr (theoretical max) |
 |
| Test throughput (Theoretical max) |
 |
 |
300 tests/hr for photometry tests only
450 tests/hr for only ISE tests
up to 300 tests/hr with mixed requests |
 |
| Number of channels (Reagent slots) |
 |
 |
42 cassette slots
3 channels on ISE module |
 |
| Programmable parameters |
 |
 |
max. 117 photometric
3 ISE tests, 8 formulas, 3 serum indices |
 |
| Sample types |
 |
 |
Serum, Plasma, Urine, CSF, Whole Blood (HbA1c only) |
 |
| Sample input (Sample disk) |
 |
 |
| Load capacity: |
55 positions on the outer ring |
| (total 110 positions) |
55 positions on the inner ring |
| |
(outer & inner ring barcode read, sample detection available) |
| STAT: |
real STAT interrupt is available |
| continuous operation: |
additional sample setting during operation |
|
 |
| Sample container types |
 |
 |
| Primary tubes |
5-10mL, 16x100, 16x75, 13x100, 13x75mm |
| Sample cup |
2.5 mL |
| Micro cup |
1,5 mL |
| Cups on tube |
Cup on top of a 16 x 75/100 mm tube
Cup on top of one non standard tube
false bottom tube |
|
 |
| Sample volume |
 |
 |
1.0 - 35 μL in 0.1 μL steps |
 |
| Sample dilution |
 |
 |
3 - 121 times, diluent > 100 μL |
 |
| Sample clot detection |
 |
 |
Available |
 |
| Minimum sample volume |
 |
 |
| Primary tubes : |
700 μL |
| Sample cup: |
100 μL |
| Micro cup: |
50 μL |
|
 |
| Sample barcode types |
 |
 |
Code 128
Codabar (NW 7)
Interleaved 2 of 5
Code 39 |
 |
| System interfaces |
 |
 |
RS 232 serial interface, bi-directional
Interface to cobas link (TeleService and automatic download) |
 |
| Sample data base |
 |
 |
10,000 routine / STAT samples |
 |
| Test methods |
 |
 |
For photometric modules: 1 point, 1 point + prozone check, 2 point, 2 point kinetic,
2 point +prozone check, 3 point, 1 point + kinetics, Rate A, Rate A + serum index,
Rate A with blank, Rate B |
 |
| Calibrator/QC Input |
 |
 |
On the sample disk, bar coded |
 |
| Calibration methods |
 |
 |
Start-up, Re-calibration
for photometric assays: Linear, non-linear multi-points, 2 point calibration, K-factor
up to 100 calibrators pre-programmable
Storage of up to 180 curves
Preventive calibration of the stand-by cassettes
Two k-factor can be defined for different sample types |
 |
| QC methods |
 |
 |
Real-time QC, individual QC, cumulative QC
up to 100 different controls pre-programmable
preventive QC after calibration of stand-by cassettes
Auto QC (without operator intervention; timer- or event- triggered) |
 |
| Rerun/reflex function |
 |
 |
Automatic rerun and manual rerun
Automatic reflex is supported by the system if reflex request is provided by a PSM or LIS |
 |
| Electrical requirements |
 |
 |
| Power requirements: |
230/400 Volts AC 3/N/PE
110 Volts AC
1.5 kVA |
| Frequency: |
50 Hz or 60 Hz +/- 0,5 % |
|
 |
| Water/waste requirements |
 |
 |
| Water: |
Bacteria free, deionised water supply:resistance of 1,0 µS/cm max. |
| Water consumption: |
Up to 12 l/hr during operation |
| Water pressure: |
0.5 - 3.5 kg / cm3, (49 - 343 Kpa) |
| Water temperature: |
15 to 30 °C |
| Biohazards waste: |
Separate container located in the system |
| L. concentrated |
Central drain port, diameter: ≥ 50 mm, |
| wall drain < waste: |
100 mm above the floor |
|
 |
| Regulatory requirements |
 |
 |
GS, CE, UL, C-UL |
 |
| Operating conditions |
 |
 |
| Ambient temperature: |
15 to 32 °C |
| Ambient humidity: |
30 to 85% (RH, without condensation) |
| Heat Output: |
2 kcal/h |
| Noise Output: |
< 65 dB |
|
 |
| Physical dimensions |
 |
 |
| Width: |
1338 mm |
| Depth: |
855 mm |
| Height: |
1262 mm |
|
 |
| Weight |
 |
 |
270 kg |
 |
 |
 |
 |
cobas e 411 analyzer - Specifications |
|
 |
| System |
 |
 |
cobas e 411 is a fully automated, random access system for immunoassay analysis.
It is available as both a disk system and a rack system. |
 |
| System components |
 |
 |
Analytical module including Window XP embedded operated touch screen PC
Sample handling module: rack or disk operated |
 |
| Sample throughput |
 |
 |
Up to 86 samples/hr |
 |
| Test throughput |
 |
 |
Up to 86 tests/hr |
 |
| Number of channels |
 |
 |
18 channels/reagent slots for up to 18 different assays |
 |
| Programmable parameters |
 |
 |
Max 60 assays definable via 2D-barcode (programming by loading) |
 |
| Sample types |
 |
 |
Serum, Plasma, Urine (others) |
 |
| Sample input/output |
 |
 |
| Load/unload capacity |
30 samples (disk)
75 samples on 15 racks |
| Rack |
RD standard 5 position rack |
| STAT handling |
Any unoccupied position on the sample disk,
dedicated STAT port on rack feeder |
|
 |
| Sample container types |
 |
 |
| Primary tubes |
5-10 mL; 16x100, 16x75, 13x100, 13x75mm |
| Sample cup |
2.5 mL |
| Micro cup |
not allowed |
| Cups on tube |
Cup on top of a 16 x 75/100 mm tube |
|
 |
| Sample volume |
 |
 |
 |
 |
10 to 50 μl per test, depending on assay protocol |
 |
| Minimum sample volume |
 |
 |
| Primary tubes: |
600 μL (13mm tube),
1,000 μL (16mm tube) |
| Sample cup: |
200 μL (standard Cup on Tube)
150 μL with special setting (reduced) |
|
 |
| Sample barcode types |
 |
 |
Code 128
Codabar (NW 7)
Interleaved 2 of 5
Code 39 |
 |
| On board control unit |
 |
 |
Tablet type PC with Pentium IV processor with coloured 15” SVGA touch screen monitor |
 |
| System interfaces |
 |
 |
RS 232 serial interface, bi-directional
Standard PC ports (USB, Ethernet, Serial etc) for other communication devices |
 |
| Sample data base |
 |
 |
2,000 tests for routine, STAT and controls |
 |
| Test methods |
 |
 |
Pre-defined assay protocols (sandwich, competitive, titration)
Heterogeneous immunoassays, competitive and sandwich (Bridging) tests 9, 18 and
27 minute tests, STAT sample capability 9 minute tests |
 |
| Calibrator/QC input |
 |
 |
Via rack or sample disk |
 |
| Calibration methods |
 |
 |
“Upon QC failure” triggered
2-point lot or cobas e pack calibration |
 |
| QC methods |
 |
 |
Individual QC + cumulative QC
Preventive QC after calibration of the stand-by bottle sets |
 |
| Electrical requirements |
 |
 |
| Power requirements: |
100-120 VAC 50/60 Hz single phase or
200-240 VAC 50/60 Hz single phase |
| Power consumption |
800 VA |
|
 |
| Water/waste requirements |
 |
 |
| Water container |
3 Liters |
| Water requirements |
> 10μS/cm or > 0.1 mega Ohm, bacteria-free |
| Water consumption |
approx. 3 L for 250 tests
approx. 12 mL/cycle |
|
 |
| Regulatory requirements |
 |
 |
CE, UL, C-UL, CB-report and certificate |
 |
| Operating conditions |
 |
 |
| Ambient temperature: |
18 to 32 °C (64.4 °F to 89.6 °F) |
| Ambient humidity: |
20% to 80% |
| Noise Output: |
60 dbA (stand-by mode)
63 dbA (avg. during operation) |
|
 |
| Physical dimensions |
 |
 |
| Width: |
1200 to 1700 mm (disk/rack) |
| Depth: |
730 to 950 mm (disk/rack) |
| Height: |
560 mm (w/o PC unit) |
|
 |
| Weight |
 |
 |
Approx. 170 kg (disk) and 210 kg (rack) |
 |
 |
|
|
 |
| |
|