Search Legal Statement
Contact Link to www.roche.com
Solutions for Diagnostics Countries







Back

 
System
Sample
Reagents
Control Unit
Technical Specs



System
System Random and continuous access, sample selective analyser
Integration of 4 measuring principles
36 tests on-board
Absorbance Photometry: Enzymes and Substrates
Turbidimetry: Specific Proteins, Drugs of Abuse
Fluorescence Polarimetry: Therapeutic Drugs, Thyroid Tests
Ion-Selective Electrode Potentiometry: Na+, K+, Cl– and Li+
Test Throughput Up to 400 tests/hr (including ISE)
Sample
Sample Types Serum, Plasma, Urine, CSF, Hemolysate and Whole Blood (HbA1c)
Sample Handling 90 primary or secondary tubes on-board
Up to 6 x RD15 racks on-board
Cooled rack position for controls and calibrators
Automatic sample dilution and concentration
Barcode reading via laser scanner, with immediate STAT recognition
Sample Container Types Primary tubes: 5 to 10ml; 16x100, 16x75, 13x100, 13x75mm
Hitachi microcup, 500µl; Hitachi standard cup, 1.5ml;
COBAS cup, 650µl; Eppendorf cup, 1.5ml; Cup on tube definable
Sample Volume Typically 2 to 10µl per test, ISE indirect 20µl, ISE direct 97µl
Sample Barcode Types Code 128, Codabar, Code 2 of 5 interleaved, Code 39
Reagents
On-board reagent capacity 32 cassettes, 50-800 tests per cassette
Up to 8 racks of 4 cassettes on-board
Automatic cassette reconstitution when required
Reagent compartment cooled to 10-15°C
On-board stability up to 3 months, calibration typically each lot
Reaction Cells Holds 1000 disposable cuvettes with 5mm path length
and 120-240µl reaction volume
Control unit
Control Unit HP workstation running Windows XP
Intel Pentium IV with 512 MB RAM
Dual 40 GB hard drives, CD ROM, floppy drive and inbuilt modem
System Interfaces RS 232 serial interface, bi-directional, modem for Remote Diagnostics access
Technical Specs
Electrical Requirements 100-125 / 200-240 Volts AC, 50 or 60Hz,
Consumption 1200VA
Physical Dimensions Width: 135cm (53.1in); Depth:66cm (25.9in); Height: 75cm (28.5in)
Weight 230kgs (506lbs)
Water Requirements Up to 2 liters per hour in operating mode, Type 1 NCCLS
Certification CE, UL, C-UL
 

Back