System |
|
 |
| System |
 |
 |
Random and continuous access, sample selective analyser
Integration of 4 measuring principles
72 tests onboard
Absorbance Photometry: Enzymes and Substrates
Turbidimetry: Specific Proteins, Drugs of Abuse
Fluorescence Polarimetry: Therapeutic Drugs, Thyroid Tests
Ion-Selective Electrode
Potentiometry: Na+,K+,Cl– and Li+ |
 |
| Test Throughput |
 |
 |
Up to 855 tests/hr (including electrolytes) |
 |
 |
 |
Sample |
|
 |
| Sample Types |
 |
 |
Serum, Plasma, Urine, CSF, Hemolysate
and Whole Blood |
 |
| Sample Handling |
 |
 |
Rack system for samples, calibrators and controls:
RD5 rack, RD15 rack, RD12+3 rack, Multi Rack; Roche standard
rack and tray compatibility; In- and Out-Tray for 75 samples each;
Automatic predilution; User definable postaction 1 & 2: dilute or
concentrate; Temperature controlled: calibrators and controls, 20°C
(68°F) Samples below 32°C (90°F); Barcode reading via LED scanner,
with immediate STAT recognition |
 |
| Sample Container Types |
 |
 |
Primary tubes: 5 to 10ml; 16x100, 16x75, 13x100, 13x75mm with
closed cap for systems with closed tube sampling option for HbA1c.
(please contact Roche for list of compatible suppliers); Hitachi
microcup, 500μl; Hitachi standard cup, 1.5ml; COBAS cup, 650μl;
Eppendorf cup, 1.5ml; Cup on tube definable |
 |
| Sample Volume |
 |
 |
Typically 2 to 10µl per test,
ISE indirect 20µl, ISE direct 97µl |
 |
| Sample Barcode Types |
 |
 |
Code 128, Codabar, Code 2 of 5 interleaved,
Code 39 |
 |
 |
 |
Reagents |
|
 |
| Reagents Onboard capacity |
 |
 |
68 cobas c packs, 50-800 tests per cobas c pack
Reagent compartment cooled to 8°C
Onboard stability up to 6 months, calibration typically each lot |
 |
| User Defined Reagents |
 |
 |
up to 5 Development Channels are available on request. |
 |
| Reaction Cells |
 |
 |
Holds 3000 disposable cuvettes with 5mm path length,
and 120-240μl reaction volume |
 |
 |
 |
Control
Unit |
|
 |
| Control Unit |
 |
 |
HP workstation with RISC processor running X Windows |
 |
| System Interfaces |
 |
 |
RS 232 serial interface, bi-directional host interface, secured digital
(web based) TeleService connection to Roche for screen sharing and
remote help |
 |
 |
 |
Technical
Specifications |
|
 |
| Electrical Requirements |
 |
 |
100-125/200-240 Volts AC, 50 or 60Hz,
consumption 1700VA |
 |
| Physical Dimensions |
 |
 |
Width: 190cm (74.8in); Depth: 90cm (35in); Height: 120 cm (47.3in) |
 |
| Weight |
 |
 |
512kg (1126lbs) |
 |
| Water Requirements |
 |
 |
5-7 liters per hour in operating mode, Type 1 NCCLS |
 |
| Certification |
 |
 |
CE, UL, C-UL |
 |
COBAS, COBAS C, COBAS INTEGRA and MODULAR are trademarks of Roche.